• Handheld Particle counters 3016IAQ\5016

    Handheld Particle counters 3016IAQ\5016


    The ergonomically designed and lightweight Lighthouse Handheld particle counters feature 0.2, 0.3 and 0.5 μm sensitivities and are the most advanced handheld particle counters on the market. Lighthouse Handheld particle counters can display cumulative and differential particle count data as well as temperature/relative humidity data on the fast and easy to read color touch screen. The Handheld can hold up to 3,000 records of particle data and store up to 50 recipes for sampling and reports. Removable batteries and an optional charger maximize uptime. Data is easily downloaded using the LMS Express software. The Handheld can be used as a mobile particle monitor or become part of a larger facility monitoring and management system.

    • Features and Advantages
    • Specifications
    • Applications

    ◆ FEATURES


    。0.2 - 25.0 μm sensitivity

    。0.1 CFM flow rate

    。Up to 6 channels of simultaneous count data

    。21 CFR Part 11 compliant secure USB flash transfer

    。Extreme life laser diode technology >20-year MTTF

    。3.5 inch color touch screen display

    。High impact injection molded plastic

    。RS-232 via RJ45 to PC or optional printer

    。3,000 record data storage

    。Reports: FS-209E, ISO-14644-1 & EU GMP Annex 1

    。2 year warranty

    3106IAQ(mass concentration) 5106
    Size Range 0.3 - 10.0μm 0.5 - 25.0μm
    Standard 6 channels

    0.2, 0.3, 0.5, 0.7, 1.0, 2.0 μm

    0.3, 0.5, 0.7, 1.0, 2.0, 5.0 μm

    0.3, 0.5, 1.0, 3.0, 5.0, 10.0 μm

    0.3, 0.5, 0.7, 1.0, 2.5, 5.0 μm

    0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm

    0.3, 0.5, 1.0. 5.0, 10.0, 25.0 μm

    Optional 6 channels

                       --

    0.5, 1.0., 3.0, 5.0, 10.0, 25.0μm

    Counting Efficiency

    50%@0.3μm;100% for particles>0.45 μm

     (per ISO 21501-4)

    50%@0.5μm;100% for particles> 0.75 μm

     (per ISO 21501-4)

    Flow rate 0.1 CFM (2.83 LPM)
    Laser Source Extreme Life Laser Diode with > 20 years MTTF
    Zero Count Level < 1 count / 5 minutes (per ISO 21501-4)
    Concentration Limits 8,000,000 particles /ft3 @ 10% coinsidence loss
    Calibration Meets ISO 21501-4 calibration using NIST traceable PSL spheres
    Count Modes Automatic, manual, beep, concentration, cumulative/differential
    Data Storage 3,000 sample records, rotating buffer, includes particle and
     environmental data, location,and time
    Communication Modes RS-232 via RJ-45 to PC or optional printer
    Supporting Software

    LMS Express, LMS Express RT, LMS Pharma, LMS Pro

    Environmental Sensors

     Direct Mount Temperature/Relative Humidity probe:
     0-150°F (-17.8 to 65.6°C)±1.8°F@77°F, 0-100% ± 5%@33%

    Touch Screen Display 3.5 inch (8.89 cm), color TFT display
    Reports FS-209E, ISO-14644-1 (1999, 2015) & EU GMP
     (Model 3016-IAQ no report standards)
    Alarms Internal, adjustable alarm on counts, low battery, sensor failure
    Sample Inlet

    Isokinetic sampling probe

    Sample Output Internally filtered to HEPA standards (>99.97% @ 0.3 μm)
    Vacuum Source Internal pump, automatic flow control
    Enclosure High impact injection molded plastic
    Power External power supply: 12 VDC, 1.25 A
    Battery

    Li-Ion, removable and rechargeable

    Dimensions 8.75“ (l) x 5.0” (w) x 2.5” (h), (22.23 x 12.70 x 6.35 cm)
    Weight 2.2 lbs (1 kg)
    Languages English (U.S.), Deutsch, Espanol, Italiano, Français, Russian, Japanese, Korean and Chinese
     

    ◆ APPLICATIONS


    • Indoor air quality study and testing

    • Searching for contamination

    • Verification and certification

    • Quality assurance