• Remote LPC Liquid Particle Counter

    Lighthouse Liquid Particle Counter Remote LPC


    LPC series products adopt the latest laser particle counting technology, elaborately designed by Lighthouse, is a professional equipment for liquid particle counting. Sensitivity from 0.1 to 100 microns, flow up to 100 ml/minute of each Lighthouse remote LPC can be real-time continuous data acquisition. This product has excellent price performance and meets the JIS standard.

    • Features and Advantages
    • Specifications
    • Applications

    ◆ FEATURES


    。Size ranges 0.1 - 100 µm

    。Up to 4 channels of data

    。100 ml per minute flow rate

    。LED status indicators for power and service

    。4-20mA or MODBUS output

    。Interfaces with facility monitoring systems

    。Compact size

    。Designed for reliability

    。2 year warranty

    Remote LPC Remote Liquid Particle Counter
    Features /  Remote LPC 0.1 µm Remote LPC 0.2 µm Remote LPC 0.3 µm Remote LPC 0.5 µm
    Size Range 0.1 - 0.5 µm 0.2 - 2.0 µm 0.3 - 3.0 µm 0.5 - 100.0 µm
    Standard 2 Channels 4-20mA 0.1, 0.2 µm 0.2, 0.3 µm 0.3, 0.5 µm 0.5, 1.0 µm
    Optional 2 Channel 4-20mA 0.1, 0.3; 0.1, 0.5 µm 0.2, 0.5; 0.2, 1.0; 0.2, 2.0 µm 0.3, 0.7; 0.3, 2.0; 0.3, 3.0 µm 0.5, 2.0; 0.5, 5.0 µm
    Standard 4 Channels Modbus 0.1, 0.2, 0.3, 0.5 µm 0.2, 0.3, 0.5, 0.7 µm 0.3, 0.5, 0.7, 1.0 µm 0.5, 0.7, 1.0, 2.0 µm
    Optional 4 Channels Modbus / 0.2, 0.5, 1.0, 2.0 µm 0.5, 1.0, 2.0, 5.0 µm
    Concentration Limit

    64,000 counts/ml @

    5% coincidence error.

    4,000 counts/ml @

    5% coincidence error

    4,000 counts/ml @

    5% coincidence error

    4,000 counts/ml @

    5% coincidence error

    Flow Rate 100 mL / minute.
    Laser Source Laser Diode
    Calibration NIST traceable.
    Data Storage Rotating buffer 2,000 records (modbus models)
    Communication Modes RS232/485 via RJ45 to PC, RS485 modbus, ASCII, Modbus or MR protocol.
    LED Indicators Power, Sampling (4-20mA model); Power, Service, Sampling (485 modbus model).
    External Alarm Output Normal open dry contact rated 0-60v AC/DC 1 Amp
    Enclosure Stainless steel.
    Sample Inlet/Outlet Connection 1/4” FlaretekTM
    Sample Temperature 32 - 140°F (0 - 60°C)
    Sample Pressure 150 PSI.
    Wetted Surface Materials Quartz, PTFE, PFA.
    Power 24 VDC
    Dimensions 5.7”(l) x 5.2”(w) x 3.0”(h) [14.47 x 13.2 x 7.6 cm].
    Weight 3.5 lbs (1.58 kg).
     

    ◆ APPLICATIONS


     wafer fab UPW monitoring

     FPD display fab UPW monitoring

     UPW station

     R&D institute