SUTO's S120 oil residues sensor (oil vapour) can monitor the residues of compressed air or gas on a long-term or sampling basis,simplely and conveniently installation, having good performance.An oil-free air compressor is hard to attain.In many industries, the monitoring of oil content is necessary in order to avoid product pollution and damage people's health.The S120 makes such monitoring affordable and reliable.The sensor provides self-calibration for better accuracy and maximum stability.The sensor's pollution level and life time will be monitored and displayed to the user.If detected beyond the range of use, the sensor blocks the sampled gas to prevent contamination.
○ Measuring the residual oil content of compressed air and gas
○ Used in both long - term and portable measurement applications
○ As low as 0.003 mg/m³ can measured
○ Quick connection through the sample tube , easy to install
○ Programmable alarm set point with device turn off option
○ Voltage/polarity display, each through output signal: - 4 ... 20 mA
- RS-485,Modbus RTU
- Relay switch (normally on)
○ Optical ion sensor, high accuracy
○ Use LED lights to indicate maintenance or alarm status
○ Can be connected to the SUTO monitor and data recorder, or to a third party monitor or controller
○ 5-inch touch screen and data recording (optional)
S120 | |
---|---|
Measure Media |
No corrosive or flammable explosive compressed air and compressed gas |
Measure range |
0.003 ... 10.00 mg/m³ (According to 1000 hPa,20°C,0% relative humidity) |
Sensor type | PID (Photoionization sensor) |
Measure lower limit | 0.003 mg/m³ |
Accuracy | 5%±0.003 mg/m³ of reading |
Operation pressure | 3 ... 15 barg (Higher pressure need customization) |
Gas relative humidity | < 40% relative humidity,no compensation |
Gas flow | < 2 l/min,Measured gas emptying into the environment |
Gas Connection | 6 mm tube |
Electrical connection | M12 connector |
Sensor life | 6000h,prompt when need replacing or maintaining |
Gas Temperature | -20° ... +50°C (entrance) |
Environment temperature | -20° ... +50°C |
Transport temperature | -30° ... +70°C |
Output singnal |
4 ... 20 mA (0 ... 10 mg/m³) RS-485,Modbus RTU 继电器开关(常开),NO, 60 VDC,1A |
Weight | 2400g |
Power | 24 VDC ± 5%,10 W |
Display and data record |
5inch touch screen,100million data records(optional) |
Shell/Size |
Polycarbonate/Aluminum alloy,271 X 205 X 91 mm |
Protection class | IP65 |
EMC | Compliance with IEC 61326-1 |
Set | Be set by SUTO or related service software with many different setting |
Sample frequency | 1s |